ARTERIOVENOUS FISTULA FAILURE DUE TO TWO STRAIGHT SHUNTS FORMATION AT DRAINING VEIN

Authors

  • YanQing Tong
  • WanLi Qi
  • Min Sun

Keywords:

Arteriovenous fistula, Shunts, Hemodialysis, Ultrasonography, Draining vein.

Abstract

A 48-year-old asian man was admitted to hospital for the consideration of hemodialysis (HD). An adequate site was decided on for the AVF after Doppler ultrasonography examination of radial artery and cephalic vein. The AVF was formed under local anaesthetic at his right wrist. When seen at the clinic appointment one month later, the AVF was not working. The subsequent Doppler ultrasonography examination revealed two straight shunts at draining vein. The failure of AVF is due to impaired outflow following reduction of the resistance between draining vein and peripheral venous system.

Author Biography

YanQing Tong

Department of Nephrology, The First Affilliated Hospital to Changchun University of Chinese Medicine, Changchun city, Gongnong Road, No 1478, Changchun city, Jilin province, China, 130021

Downloads

Published

2013-03-18

How to Cite

Tong, Y., Qi, W., & Sun, M. (2013). ARTERIOVENOUS FISTULA FAILURE DUE TO TWO STRAIGHT SHUNTS FORMATION AT DRAINING VEIN. African Journal of Traditional, Complementary and Alternative Medicines, 10(3), 541–542. Retrieved from https://athmsi.org/journals/index.php/ajtcam/article/view/1804